HW group Temp-1W-Pt100

SPHINX in Europa
DE
  • AT
  • BE
  • DK
  • FR
  • DE
  • IE
  • IT
  • NL
  • ES
  • SE
  • CH
  • UK
Produktgruppen > Environmental Sensors > HW group Temp-1W-Pt100

HW group Temp-1W-Pt100

Double converter to connect two Pt-100 and Pt-1000 probes to 1-Wire UNI (RJ11)

HW group Temp-1W-Pt100
5122_16401.png
5122_11895.jpg
5122_11896.png
5122_11897.png
  • EigenschaftenEigenschaften
  • BeschreibungBeschreibung
  • Weitere InformationenWeitere Informationen
  • PDFPDF
  • Up to two external temperature probes (Pt100 or Pt1000) can be connected
  • Measured temperature range is -200 to +850 °C (-328 to 1560 °F)
  • Supports 2/3/4-wire probe connection
  • High temperature measurement accuracy, auto-range
  • Sensors can be daisy-chained (two RJ11 connectors)
  • Converter can be calibrated at 8 points
  • Calibration protocols with a particular temperature sensor can be provided for an extra fee
  • Powered from the 1-Wire UNI bus (RJ11)

Double converter to connect two Pt-100 and Pt-1000 probes to 1-Wire UNI (RJ11). One or two external temperature probes can be monitored over LAN (Poseidon2) and GSM (Ares). Typical applications include measuring extreme temperatures in furnaces, coolers, engines, etc.

The converter connects one or two Pt100 or Pt1000 platinum temperature sensors to the 1-Wire UNI bus. Pt100 probes are available in numerous form factors designed for various uses.

The “Converter 2xPt100 1W-UNI” can be connected to Poseidon2 or Ares units. The values can be shared over LAN or GSM.

  • The converter automatically detects Pt100 and Pt1000 probes
  • Measured temperature range is -200 to +850 °C (-328 to 1560 °F)
  • The external probes can be connected using 2, 3, or 4 wires
  • The converter is powered from the 1-Wire UNI bus (RJ11)
  • High temperature measurement accuracy, auto-range
  • Sensors can be daisy-chained (two RJ11 connectors)
  • Converter can be calibrated at 8 points



Applications

  • Accurate measurement of temperature in a wide range
  • Cooling plants with built-in temperature sensors at fixed locations
  • Large-scale measurement systems with many measuring points
  • Measuring HVAC and environment conditions in smart buildings
  • Measurement of temperature in food and crop storages
  • Industrial systems involving accurate temperature measurement
  • Temperature monitoring in pharmaceutical storages


ProdGrafik

Bus
Type 1Wire-UNI (RJ11)
Connects to Poseidon2 xxxx, Ares xx, Poseidon 2250, Poseidon 4002, SiteMon5
Connector 2x RJ11 (sensors can be daisy-chained)
Maximum distance 60m (total line length from the active port)
Electrical isolation 1Wire-UNI bus is NOT electrically isolated from the temperature probes
Power
Power supply 1Wire-UNI bus
Power limits One active RJ11 port can power at most two converters.
To boost the power, use our “1-Wire hub Power”.
Converter input
Connector 2x 4-pin terminal blocks (1,2,3,4 + 5,6,7,8)
Supported sensors 2x Pt100 or Pt1000 temperature probe (autodetected)
Probe connection 2-wire, 3-wire or 4-wire connection
Input range / accuracy Pt-100:-200°C to +650°C (± 0.3°C at +25°C converter temperature)
Pt-100:-200°C to +850°C (± 0.6°C at +25°C converter temperature)

Pt-1000: -120°C to +650°C (± 0.3°C at +25°C converter temperature)
Pt-1000: -120°C to +850°C (± 0.6°C at +25°C converter temperature)
Measurement system accuracy The total precision is the converter precision + Pt-100 probe precision.
For example, for Pt100/A: Ät = ±(0.15 + 0.002 x |t|) [°C]

Overall accuracy = ±(0.3 + 0.1 + 0.15 + 0.002 x |t|) [°C] (at +25°C converter temperature)

Overall accuracy = ±0.55 °C when measuring 0 °C (converter +25°C)
Overall accuracy = ±0.75 °C when measuring 100 °C (converter +25°C)
Overall accuracy = ±2.25 °C when measuring 700 °C (converter +25°C)
Sensor output
Resolution 1 decimal place
Displayed units Default [°C]
For conversion / recalculation to other units, please contact your distributor.
Measuring period Once per second
Miscellaneous
Operating conditions -40 to +85 °C (-40 to +185 °F) / 5 to 70% RH
Dimensions / mass 65 x 80 x 30 [mm] / 200 g



Temperature measurement with Pt100 and Pt1000 over the Ethernet

  • A Pt100 probe is connected to the "Converter 2xPt100 1W-UNI" using 3 wires.
  • A Pt1000 probe is connected to the "Converter 2xPt100 1W-UNI" using 2 wires.
  • The converter is connected to a Poseidon2 3266 remote monitoring LAN unit over the 1-Wire UNI (RJ11) bus.
  • "Converter 2xPt100 1W-UNI" counts as two sensors in the Poseidon2 3266 unit.
  • The converter is powered from the 1-Wire UNI bus (RJ11). Additional sensors can be daisy-chained to this sensor.

  • Poseidon2 3266 displays the Pt100 and Pt1000 temperatures in its web interface. Temperature units – [°C] / [°K] / [°F] – can be configured.
  • Poseidon2 3266 sends an e-mail or a text message (SMS) when a temperature is outside of a specified "SafeRange".
  • Poseidon2 3266 shares data over the LAN using SNMP, XML (http) and Modbus/TCP.
  • Poseidon2 3266 can be connected to the SensDesk.com online portal.
  • Sensor readings can be displayed using an Android or iOS (Apple) app for tablets or mobile phones.



Temperature measurement with Pt1000 over GSM

  • A Pt1000 probe is connected to the "Converter 2xPt100 1W-UNI" using 2 wires.
  • "Converter 2xPt100 1W-UNI" is connected to an Ares 12 – GSM remote measurement unit.
  • The converter is powered from the 1-Wire UNI bus (RJ11). Additional sensors can be daisy-chained to this sensor.
  • An additional outdoor temperature and humidity sensor is connected to the Ares unit, too.
  • The internal Ares battery provides back-up power to both sensors.

  • Ares 12 is a GSM/GPRS unit for remote measurement and monitoring.
  • Ares 12 sends an e-mail or a text message (SMS) when a temperature is outside of a specified "SafeRange".
  • Temperature units – [°C] / [°K] / [°F] – can be configured in the Ares 12.
  • Ares 12 can be connected to the SensDesk.com online portal.
  • Sensor readings can be displayed using an Android or iOS (Apple) app for tablets or mobile phones.



Ordering

Temp-1W-Pt100 Cable Pt-100 based temperature sensor with 1-Wire UNI output.
Contains a Pt-100 probe (–50 to 200°C, Pt100/A, 2m silicon cable) and a 1-Wire UNI converter (Converter 2xPt100 1W-UNI).
Second Pt-100 or Pt-1000 temperature probe can be connected.
Temp-1W-UNI Pt100 Frost Pt-100 based low temperatures sensor with 1-Wire UNI output.
Contains a Pt-100 temperature probe (-190 to +150°C, PTFE covered cable 2m,Pt100/B) and a 1-Wire UNI converter (Converter 2xPt100 1W-UNI).
Second Pt-100 or Pt-1000 temperature probe can be connected.
Converter 2xPt100 1W-UNI Converter for connecting 2x Pt100 or Pt100 temperature probes to 1-Wire UNI bus (RJ11).
External temperature probe is not included in the package.
Probe Pt100 TG8 2m Temperature probe Pt100/A on the cable.
Temperature range -50 to 200°C.
Silicon covered cable 2m, 4-wires, IP67 probe in in stainless steel tube 50mm.
Probe Pt100 TR125 2m Temperature probe Pt100/B for low temperatures.
Temperature range -190 to +150°C.
PTFE covered cable 2m, 4-wires, IP67 probe in in stainless steel tube 50mm.
   More accessories
1-Wire hub Power Hub for connecting 1 to 8 sensors (RJ11), with power boost.
External 12VDC power adapter can be connected.
Suitable for 1-Wire and 1-Wire UNI bus.
T-Box RJ11 Hub for connecting 1 to 5 sensors (5x RJ11), 10 cm connecting cable.
T-Box2 Hub for 1 or 2 sensors (2x RJ11) – 3m connecting cable.



Kunden, die diesen Artikel gesehen haben, sahen auch
Alle Cookies akzeptieren Ablehnen Konfigurieren

Diese Website verwendet Cookies, um eine bestmögliche Erfahrung bieten zu können. Mehr erfahren.